User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Omega DP-420:Transducer Indicator DATE: 09-Jun-95 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 19 NUMBER OF LINES: 181 STANDARD: Measurements Group 1550A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F 1.002 ASK+ X 1.003 HEAD ÍÍ INITIAL CONDITIONS ÍÍ 1.004 DISP The following Standard test equipment is required: 1.004 DISP 1. Measurements Group 1550A Strain Indicator Calibrator 1.004 DISP ÿ DTB # 55-2 1.004 DISP 2. Strain Indicator Calibration Cable 1.005 STD Measurements Group 1550A 1.006 HEAD {} 1.007 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.008 HEAD {º LINEARITY TEST º} 1.009 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.010 HEAD {} 1.011 HEAD {Full scale = 9000} 1.012 HEAD {}ÍÍ LINEARITY TEST ÍÍ 1.013 JMP 2.001 1.014 EVAL 2.001 DISP With the UUT disconnected from AC power, remove the 2.001 DISP UUT front panel. 2.002 PIC 420 (22,2,W/B) "Set the EXCITATION switch to the 10V" 2.002 PIC (23,2,W/B) "position. " 2.003 DISP Set the 1550A 350ê bridge S+ shorting link to the closed 2.003 DISP position. 2.004 DISP Set the 1550A front panel controls as follows: 2.004 DISP MICROSTRAIN buttons ................................ 000 2.004 DISP CIRCUIT button .................................... 350ê 2.004 DISP POLARITY button ........................... positive (+) 2.005 DISP Connect the Strain Indicator Calibration cable to the 2.005 DISP UUT rear panel connector. 2.006 DISP ÿ Connect the 1550A and the UUT using the Strain 2.006 DISP ÿ Indicator Calibration cable as follows: 2.006 DISP ÿ [27][91]1m 1550A TO PLDT 2.006 DISP ÿ P+ ÄÄÄÄÄÄÄÄÄ> P+ (green & white) 2.006 DISP ÿ P- ÄÄÄÄÄÄÄÄÄ> P- (white) 2.006 DISP ÿ S+ ÄÄÄÄÄÄÄÄÄ> S+ (orange & red) 2.006 DISP ÿ S- ÄÄÄÄÄÄÄÄÄ> S- (black,blue & GND) 2.007 DISP Connect the UUT to AC power. 2.008 DISP Adjust the 1550A 350ê ñ ZERO control, while togling 2.008 DISP between the 1550A + and - POLARITY switch positions, 2.008 DISP for a UUT reading that is the same in magnitude for 2.008 DISP both positions. 2.009 DISP Set the 1550A front panel controls as follows: 2.009 DISP POLARITY button ........................... positive (+) 2.010 PIC 420 (22,39,W/B) "Adjust the FINE TARE (ZERO) and " 2.010 PIC (23,39,W/B) "COARSE TARE for a UUT display of 00." 2.011 DISP Set the 1550A front panel controls as follows: 2.011 DISP MICROSTRAIN PER STEP buttons ....................... 090 2.012 PIC 420 (22,1,W/B) "Set SPAN switch settings and adjust " 2.012 PIC (23,1,W/B) "FINE SPAN for a display of 9000. " 2.013 DISP Set the 1550A front panel controls as follows: 2.013 DISP MICROSTRAIN PER STEP buttons ....................... 000 2.014 PIC 420 (22,39,W/B) "Adjust the FINE TARE (ZERO) and " 2.014 PIC (23,39,W/B) "COARSE TARE for a UUT display of 00." 2.015 DISP Set the 1550A front panel controls as follows: 2.015 DISP MICROSTRAIN PER STEP buttons ....................... 001 2.016 PIC 420 (22,29,W/B) "Adjust ZERO GAP for a UUT display " 2.016 PIC (23,29,W/B) "of 100. " 2.017 DISP Set the 1550A front panel controls as follows: 2.017 DISP MICROSTRAIN PER STEP buttons ....................... 090 2.018 OPBR Does the UUT read 9000 (full scale)? 2.019 JMPF 2.012 2.020 DISP Set the 1550A front panel controls as follows: 2.020 DISP MICROSTRAIN PER STEP buttons ....................... 080 2.021 MEMI Use keyboard to enter actual UUT reading: 2.022 MATH MEM1 = 8000 2.023 MEME 2.024 MEMC æî 9U 3.001 DISP Set the 1550A front panel controls as follows: 3.001 DISP MICROSTRAIN PER STEP buttons ....................... 070 3.002 MEMI Use keyboard to enter actual UUT reading: 3.003 MATH MEM1 = 7000 3.004 MEME 3.005 MEMC æî 9U 4.001 DISP Set the 1550A front panel controls as follows: 4.001 DISP MICROSTRAIN PER STEP buttons ....................... 060 4.002 MEMI Use keyboard to enter actual UUT reading: 4.003 MATH MEM1 = 6000 4.004 MEME 4.005 MEMC æî 9U 5.001 DISP Set the 1550A front panel controls as follows: 5.001 DISP MICROSTRAIN PER STEP buttons ....................... 050 5.002 MEMI Use keyboard to enter actual UUT reading: 5.003 MATH MEM1 = 5000 5.004 MEME 5.005 MEMC æî 9U 6.001 DISP Set the 1550A front panel controls as follows: 6.001 DISP MICROSTRAIN PER STEP buttons ....................... 040 6.002 MEMI Use keyboard to enter actual UUT reading: 6.003 MATH MEM1 = 4000 6.004 MEME 6.005 MEMC æî 9U 7.001 DISP Set the 1550A front panel controls as follows: 7.001 DISP MICROSTRAIN PER STEP buttons ....................... 030 7.002 MEMI Use keyboard to enter actual UUT reading: 7.003 MATH MEM1 = 3000 7.004 MEME 7.005 MEMC æî 9U 8.001 DISP Set the 1550A front panel controls as follows: 8.001 DISP MICROSTRAIN PER STEP buttons ....................... 020 8.002 MEMI Use keyboard to enter actual UUT reading: 8.003 MATH MEM1 = 2000 8.004 MEME 8.005 MEMC æî 9U 9.001 DISP Set the 1550A front panel controls as follows: 9.001 DISP MICROSTRAIN PER STEP buttons ....................... 010 9.002 MEMI Use keyboard to enter actual UUT reading: 9.003 MATH MEM1 = 1000 9.004 MEME 9.005 MEMC æî 9U 10.001 DISP Set the 1550A front panel controls as follows: 10.001 DISP MICROSTRAIN PER STEP buttons ....................... 009 10.002 MEMI Use keyboard to enter actual UUT reading: 10.003 MATH MEM1 = 900 10.004 MEME 10.005 MEMC æî 9U 11.001 DISP Set the 1550A front panel controls as follows: 11.001 DISP MICROSTRAIN PER STEP buttons ....................... 008 11.002 MEMI Use keyboard to enter actual UUT reading: 11.003 MATH MEM1 = 800 11.004 MEME 11.005 MEMC æî 9U 12.001 DISP Set the 1550A front panel controls as follows: 12.001 DISP MICROSTRAIN PER STEP buttons ....................... 007 12.002 MEMI Use keyboard to enter actual UUT reading: 12.003 MATH MEM1 = 700 12.004 MEME 12.005 MEMC æî 9U 13.001 DISP Set the 1550A front panel controls as follows: 13.001 DISP MICROSTRAIN PER STEP buttons ....................... 006 13.002 MEMI Use keyboard to enter actual UUT reading: 13.003 MATH MEM1 = 600 13.004 MEME 13.005 MEMC æî 9U 14.001 DISP Set the 1550A front panel controls as follows: 14.001 DISP MICROSTRAIN PER STEP buttons ....................... 005 14.002 MEMI Use keyboard to enter actual UUT reading: 14.003 MATH MEM1 = 500 14.004 MEME 14.005 MEMC æî 9U 15.001 DISP Set the 1550A front panel controls as follows: 15.001 DISP MICROSTRAIN PER STEP buttons ....................... 004 15.002 MEMI Use keyboard to enter actual UUT reading: 15.003 MATH MEM1 = 400 15.004 MEME 15.005 MEMC æî 9U 16.001 DISP Set the 1550A front panel controls as follows: 16.001 DISP MICROSTRAIN PER STEP buttons ....................... 003 16.002 MEMI Use keyboard to enter actual UUT reading: 16.003 MATH MEM1 = 300 16.004 MEME 16.005 MEMC æî 9U 17.001 DISP Set the 1550A front panel controls as follows: 17.001 DISP MICROSTRAIN PER STEP buttons ....................... 002 17.002 MEMI Use keyboard to enter actual UUT reading: 17.003 MATH MEM1 = 200 17.004 MEME 17.005 MEMC æî 9U 18.001 DISP Set the 1550A front panel controls as follows: 18.001 DISP MICROSTRAIN PER STEP buttons ....................... 001 18.002 MEMI Use keyboard to enter actual UUT reading: 18.003 MATH MEM1 = 100 18.004 MEME 18.005 MEMC æî 9U 19.001 DISP Replace UUT front panel. 19.002 HEAD {} 19.003 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 19.003 DISP {º THIS COMPLETES THE VERIFICATION º} 19.003 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 19.004 END